Catalog Number 5100009000 |
Device Problem
Mechanical Problem (1384)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 01/29/2015 |
Event Type
malfunction
|
Event Description
|
The universal handswitch was returned for service.Upon evaluation at the manufacturer, it was discovered that the run/safe switch would not lock into either position, creating a situation from which the device may activated unintentionally.There was no patient involvement, and there were no user injuries or adverse consequences.
|
|
Event Description
|
The universal handswitch was returned for service.Upon evaluation at the manufacturer, it was discovered that the run/safe switch would not lock into either position, creating a situation from which the device may activated unintentionally.There was no patient involvement, and there were no user injuries or adverse consequences.
|
|
Manufacturer Narrative
|
A follow up report will be filed after the quality investigation has been completed.
|
|
Manufacturer Narrative
|
The reported event was duplicated.It was confirmed by the service technician through visual inspection the run/safe switch was not locking into place.Impact damage may cause or contribute to the reported event.The handswitch is not a repairable device and will not be returned to the user facility.
|
|