MAUDE Adverse Event Report: BECKMAN COULTER COULTER LH 780 HEMATOLOGY ANALYZER; COUNTER, DIFFERENTIAL CELL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECKMAN COULTER COULTER LH 780 HEMATOLOGY ANALYZER; COUNTER, DIFFERENTIAL CELL Back to Search Results
Catalog Number 723585
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/28/2015
Event Type  malfunction  
Event Description
While troubleshooting a different issue the field service engineer (fse) found that the hemoglobin amplifier of the coulter lh 780 hematology analyzer was broken.Erroneous patient results were not generated and there was no change or affect to patient treatment in connection with this event.
 
Manufacturer Narrative
A field service engineer (fse) evaluated the instrument on (b)(4) 2015.While troubleshooting a reported leak, the fse found that the hgb preamplifier was broken.The fse replaced the preamplifier, which resolved the issue.The repairs were verified per established procedures.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COULTER LH 780 HEMATOLOGY ANALYZER
Type of Device
COUNTER, DIFFERENTIAL CELL
Manufacturer (Section D)
BECKMAN COULTER
250 south kraemer blvd.
brea CA 92821
Manufacturer (Section G)
BECKMAN COULTER
11800 sw 147th ave
miami FL 33196
Manufacturer Contact
rula melconian
250 s. kraemer blvd.
brea, CA 92821-8000
7149614295
MDR Report Key4552513
MDR Text Key5461105
Report Number1061932-2015-00316
Device Sequence Number1
Product Code GKZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K061616
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/28/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/26/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number723585
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/28/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/01/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-