• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECKMAN COULTER COULTER LH 750 HEMATOLOGY ANALYZER; COUNTER, DIFFERENTIAL CELL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECKMAN COULTER COULTER LH 750 HEMATOLOGY ANALYZER; COUNTER, DIFFERENTIAL CELL Back to Search Results
Catalog Number 6605632
Device Problem Device Emits Odor (1425)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/30/2015
Event Type  malfunction  
Event Description
The customer reported a burning smell and a diluter 2 card failure from a coulter lh 750 hematology analyzer.The customer did not see any sparks, arcs, of flames, and the fire department was not called.The customer was not exposed or injured, and no emergency or response procedures were initiated.Erroneous patient results were not generated and there was no change or affect to patient treatment in connection with this event.
 
Manufacturer Narrative
A field service engineer (fse) evaluated the instrument on 01/30/2015.The fse indicated that he did not observe burnt components.The fse performed multiple troubleshooting procedures, including diluter 2 and interconnect board replacements; however, the issue was not resolved.He replaced the diluter solenoid optics power supply board on the left door to resolve the issue.The repairs were verified per established service procedures.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COULTER LH 750 HEMATOLOGY ANALYZER
Type of Device
COUNTER, DIFFERENTIAL CELL
Manufacturer (Section D)
BECKMAN COULTER
250 s kraemer blvd
brea CA 92821
Manufacturer (Section G)
BECKMAN COULTER
11800 sw 147th ave
miami FL 33196
Manufacturer Contact
rula melconian
250 s. kraemer blvd.
brea, CA 92821
7149614295
MDR Report Key4555639
MDR Text Key5456562
Report Number1061932-2015-00329
Device Sequence Number1
Product Code GKZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K061574
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/30/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/28/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number6605632
Other Device ID NumberSOFTWARE VERSION: 2D1
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/30/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/01/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-