The arterial line pressure is set to stop at 360mmhg on the heart lung machine (hlm) pressure module.By the time the pump actually stops the transducer pressure is measuring at 670 - 700mmhg, and can go even higher during a case when the pump's rpm is set even higher.There is an inherent delay in response time from when the hlm senses the overpressure and then when the pump actually stops.Setting the occlusion rate can be frustrating because as the clinician changes the occlusion it can take several seconds for the reading to respond.Likewise, when the arterial alarm limit is exceeded it takes a moment for the pump to stop, which allows for the over pressure.This response happens 100% of the time and is entirely predictable.We purchased the new s5 systems last spring and perfusionists state the older s3 systems had a much better reaction time.The issue has been addressed with the sorin sales representative and also the regional service manager but there has been no support from sorin on the problem.We have seen one instance of an occluded arterial line bursting from over pressure; fortunately the patient was not adversely affected.This happens 100% of cases (nothing about the specific patient).
|