ST. JUDE MEDICAL, INC., CRMD CONFIRM; IMPLANTABLE CARDIAC MONITOR Back to Search Results
Model Number DM2102
Device Problems Failure to Capture (1081); Failure to Sense (1559)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 11/27/2014
Event Type  malfunction  
Event Description
It was reported that during implant procedure, the implantable cardiac monitor exhibited loss of capture.The physician attempted a different position but was unsuccessful.The device was no longer used and replaced.No patient symptoms were reported.
 
Manufacturer Narrative
All information provided by manufacturer, no medwatch form was received.Device evaluation anticipated, but not yet begun.
 
Manufacturer Narrative
All information provided by manufacturer, no medwatch form was received.(b)(4).Final analysis found an anomalous hybrid which resulted in a sensing anomaly.
 
Event Description
Corrected information the implantable cardiac monitor exhibited loss of sensing and not loss of capture.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONFIRM
Type of Device
IMPLANTABLE CARDIAC MONITOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC., CRMD
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC., CRMD
15900 valley view court
sylmar CA 91342
Manufacturer Contact
jennie hamel
15900 valley view court
sylmar, CA 91342
8184933462
MDR Report Key4560290
MDR Text Key20367258
Report Number2017865-2015-02373
Device Sequence Number1
Product Code MXC
Combination Product (y/n)N
PMA/PMN Number
K081365
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,foreign,health profe
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/02/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2015
Device Model NumberDM2102
Device Catalogue NumberSMTFY999
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/20/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received01/07/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/18/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-