No injuries or procedural delays or cancellations were reported in association to the event.The procedure was successfully completed, without delay as the facility has 3 lighting systems and continued the procedure with the 2 remaining functional lightheads.A steris service technician arrived at the facility, inspected the lighting system and found the lamp head power connector exhibited signs of overheating.The technician replaced the lamp holder, tested the unit and confirmed the lighthead was operating according to specification.The lighting system was installed in february of 2004 and is not under steris service contract.The facility is responsible for ensuring the lighting system is properly maintained and receives the recommended preventive maintenance as outlined in the harmony la 500 dual surgical light operator manual.Section 6.2 of the operator manual states, "2.Inspect the lamps for damage or discoloration.Be sure the filament of the lamp in the secondary position is intact.Replace if necessary." no further issues have been reported.
|