• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECKMAN COULTER COULTER LH 750 HEMATOLOGY ANALYZER; COUNTER, DIFFERENTIAL CELL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECKMAN COULTER COULTER LH 750 HEMATOLOGY ANALYZER; COUNTER, DIFFERENTIAL CELL Back to Search Results
Catalog Number 6605632
Device Problems Thermal Decomposition of Device (1071); Device Emits Odor (1425)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/30/2015
Event Type  malfunction  
Event Description
While servicing a coulter lh 750 hematology analyzer at a customer site, the field service engineer (fse) observed a burn smell from the instrument and a burn mark on the power supply cable.There was no report of smoke, flames or arcing; a fire extinguisher was not used and the fire department was not called in connection with this event.There was no injury associated with the event.
 
Manufacturer Narrative
The fse discovered a loose t-fitting inside the power supply for the 60 psi pressure.The pneumatic power supply was replaced, resolving this issue.The fse replaced the power supply cable that connects the wall outlet to the power supply, resolving this issue.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COULTER LH 750 HEMATOLOGY ANALYZER
Type of Device
COUNTER, DIFFERENTIAL CELL
Manufacturer (Section D)
BECKMAN COULTER
250 s kraemer blvd
brea CA 92821
Manufacturer (Section G)
BECKMAN COULTER
11800 sw 147th ave
miami FL 33196
Manufacturer Contact
miranda holland
11800 sw 147th avenue, 32-l23
miami, FL 33196
3053802031
MDR Report Key4573275
MDR Text Key5559009
Report Number1061932-2015-00387
Device Sequence Number1
Product Code GKZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K061574
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 01/30/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/05/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number6605632
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/30/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/01/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-