The customer reported to maquet sas that during position of the lighthead, some metal dust, at the junction between the fork and the spring arm, fell down on to the surgical field.No injuries were reported.(b)(4).
|
A maquet tech evaluated that the cause of the metal dust is related to a lack of grease at the junction between the fork and the spring arm, leading to a grinding during manipulation of the lighthead.Maquet has not been informed of any maintenance being done on this device.In the powerled operating manual, maquet suggests to customers that the light should be removed and lubricated at the sleeve every year by a certified tech.The tech replaced the fork, added grease, and returned the device to service.
|