• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NAVILYST MEDICAL NAVILYST MEDICAL; STOPCOCK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NAVILYST MEDICAL NAVILYST MEDICAL; STOPCOCK Back to Search Results
Catalog Number H965700550081
Device Problem Hole In Material (1293)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/30/2015
Event Type  malfunction  
Event Description
As reported by navilyst medical's distributor in (b)(4), in the distributor's warehouse, a small hole was found in the tyvek portion of the stopcock pouch, breaching the sterility.The device had not been provided to a hospital and was returned to navilyst medical for evaluation.
 
Manufacturer Narrative
A review of the device history records was performed for the reported packaging lot for any deviations related to the reported defect of the complaint.The review confirms that the lot met all material, assembly and performance specifications.The navilyst medical (b)(6) 2015 complaint report was reviewed for the product family of stopcocks/manifolds and the failure mode, "hole in pouch." no adverse trends were identified.The returned sample was visually inspected and the hole in the pouch was confirmed.As part of the receiving process at nipro (the distributor in (b)(4)), all pouched products are removed from their inner boxes and a nipro label (in (b)(4)) is applied to each pouch.Additional handling may occur if the pouches are 100% visually inspected.The pouched product is then re-boxed into the inner box by the nipro warehouse employees.The hole in the tyvek was likely caused by the handle of the stopcock, however, what caused the handle to be pushed against the tyvek in a manner that resulted in a hole cannot be determined.Potential contributing factors include: over-crowding of filled pouches in the in-process bins or improper handling of pouches as they are placed in inner boxes at navilyst medical.Handling during transit to the nipro warehouse.Handling at nipro during the labeling/inspection and re-boxing process.All pouches are 100% inspected per navilyst medical procedures during the sealing and final box processes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVILYST MEDICAL
Type of Device
STOPCOCK
Manufacturer (Section D)
NAVILYST MEDICAL
glens falls NY 12801
Manufacturer (Section G)
NAVILYST MEDICAL
10 glens falls technical park
glens falls NY 12801
Manufacturer Contact
michael duerr
10 glens falls technical park
glens falls, NY 12801
5187424571
MDR Report Key4577633
MDR Text Key5546535
Report Number1317056-2015-00066
Device Sequence Number1
Product Code FMG
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K842829
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Distributor
Reporter Occupation Not Applicable
Type of Report Initial
Report Date 02/06/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/27/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date12/31/2017
Device Catalogue NumberH965700550081
Device Lot Number4829985
Other Device ID NumberSTOPCOCK
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/17/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received02/06/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/01/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-