As reported by navilyst medical's distributor in (b)(4), in the distributor's warehouse, a small hole was found in the tyvek portion of the stopcock pouch, breaching the sterility.The device had not been provided to a hospital and was returned to navilyst medical for evaluation.
|
A review of the device history records was performed for the reported packaging lot for any deviations related to the reported defect of the complaint.The review confirms that the lot met all material, assembly and performance specifications.The navilyst medical (b)(6) 2015 complaint report was reviewed for the product family of stopcocks/manifolds and the failure mode, "hole in pouch." no adverse trends were identified.The returned sample was visually inspected and the hole in the pouch was confirmed.As part of the receiving process at nipro (the distributor in (b)(4)), all pouched products are removed from their inner boxes and a nipro label (in (b)(4)) is applied to each pouch.Additional handling may occur if the pouches are 100% visually inspected.The pouched product is then re-boxed into the inner box by the nipro warehouse employees.The hole in the tyvek was likely caused by the handle of the stopcock, however, what caused the handle to be pushed against the tyvek in a manner that resulted in a hole cannot be determined.Potential contributing factors include: over-crowding of filled pouches in the in-process bins or improper handling of pouches as they are placed in inner boxes at navilyst medical.Handling during transit to the nipro warehouse.Handling at nipro during the labeling/inspection and re-boxing process.All pouches are 100% inspected per navilyst medical procedures during the sealing and final box processes.
|