Model Number 3608 |
Device Problem
Premature End-of-Life Indicator (1480)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
Injury
|
Manufacturer Narrative
|
Results: pulse load testing determined the device battery is depleted to the end of life.An accurate longevity could not be performed as the programming history and lead impedance history was unknown.The observation of an ipg low battery message was confirmed.The device functioned within specification throughout analysis.Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
|
|
Event Description
|
It was reported the patient ((b)(6)) received a low battery warning due to possible premature battery depletion, however, the patient did not lose stimulation.Surgical intervention was undertaken and the ipg was explanted and replaced.The patient reported effective stimulation coverage postoperative.
|
|
Manufacturer Narrative
|
(b)(4).Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
|
|
Search Alerts/Recalls
|