Model Number BI-700-00027-120
Device Problems Device Emits Odor (1425); Device Issue (2379)
Patient Problem No Patient Involvement (2645)
Event Date 02/12/2015
Event Type  malfunction  
Event Description
A site representative reported that the imaging system was charging and there was a burning/chemical smell coming from the system.The site representative unplugged the system from the wall to stop battery charging.There was no patient present when this issue was identified.
 
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.A medtronic representative, following up with the site, tested the system and found 0 volts on pins 5, 6, and 7 of junction 7 indicating bad batteries in tray 4.Replacement batteries shipped to site for issue resoluition.The medtronic representative replaced the batteries in tray 4, performed gain calibrations and reported the system is ready to use.A medtronic representative performed am imaging system system check-out, all areas passed, after the batteries were replaced.No parts have been received by the manufacturer for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-ARM 1000 IMAGING SYSTEM
Type of Device
IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer Contact
kristianna bilan
826 coal creek circle
louisville, CO 80027-9710
7208902338
MDR Report Key4596428
MDR Text Key5451601
Report Number1723170-2015-00301
Device Sequence Number1
Product Code OXO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050996
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Health Professional
Type of Report Initial
Report Date 02/12/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/12/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberBI-700-00027-120
Device Catalogue Number9732719
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/12/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/10/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-