It was reported that during use of the device for a cardiopulmonary bypass procedure, the heater cooler unit would not cool.There was a delay in the middle cardioplegia (cpg) delivery of about ten to twelve minutes.The customer turned the unit on high flush, switched the two sides, but did not help the issue.As a result, an alternate device was employed.The surgical procedure was completed successfully.There was no blood loss, nor adverse consequences to the patient.Per the clinical review on (b)(6) 2015: during the second dose of blood cardioplegia, the unit did not cool the cardioplegia solution to the desired temperature level.According to the perfusionist (ccp), the low water level light emitting diode (led) illuminated, even though the water level appeared to be adequate.The dose of cpg was stopped and a second water device (cincinnati sub zero unit) was brought into the operating room (o/r) and used for the cpg circuit.The second dose of cpg was completed once the additional water system was ready.According to the ccp, this delayed the procedure by about ten minutes.The heater cooler unit continued to be used for the oxygenator water supply.According to the ccp, the case was completed successfully without associated blood loss.The patient was weaned from cpb in normal fashion and there was no harm observed.
|