On (b)(6) 2015, zefon received a customer complaint where the customer stated that pressure infusor device used for rapid fluid infusion had lost pressure during use.No patient harm was reported.The subject device was returned for evaluation on (b)(6) 2015 and examination concluded that the device held pressure and no leaks were detected.Based on the device hazard analysis, loss of pressure would result in fluids not being infused at the intended rate (i.E, slower infusion rate, through gravity forces only).The risk of significant injury to the patient in this circumstance has been determined to be very low.On (b)(6) 2015, the findings were discussed with the clinical nurse specialist at the hospital.At this time, zefon became aware that the complainant hospital was using the device for an off label purpose.The device is marketed and labeled as an infusor for rapid fluid infusion, but the hospital was using the device for invasive arterial bp pressure monitoring.This off label use requires the pressure cuff to be inflated and maintained under pressure for longer periods of time than is required for rapid fluid infusion.
|