The referenced clv-s190 was returned to the olympus medical systems corporation (omsc) for eval.The eval could not confirm the user's report, however the dust of gauze was found on the aperture unit.It would appear that this phenomenon was attributed to temporary electrical loss for the iris unit because of the dust.Omsc stated the iris handling of the clv-s190 in the instruction manual when the clv-s190 had abnormalities.There were no further details provided.If significant additional info is received, this report will be supplemented.This report is being submitted as a medical device report in an abundance of caution.
|