• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - GALWAY CAROTID WALLSTENT?; STENT, CAROTID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - GALWAY CAROTID WALLSTENT?; STENT, CAROTID Back to Search Results
Model Number H7493915010240
Device Problems Bent (1059); Kinked (1339)
Patient Problem No Patient Involvement (2645)
Event Date 02/03/2015
Event Type  malfunction  
Event Description
It was reported that stent damage occurred.A 10.0-24 carotid wallstent¿ was selected for use to treat the lesion.During unpacking, the physician noticed that the device was slightly kinked approximately 40 degrees at the stent mounted position.No patient complications were reported and the patient's status was good.
 
Manufacturer Narrative
Age at time of event: 18 years or older.(b)(4).Device evaluated by mfr: the device was not returned for evaluation.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is considered handling damage as the event occurred without direct patient contact.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAROTID WALLSTENT?
Type of Device
STENT, CAROTID
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
ballybrit business park
galway
Manufacturer (Section G)
BOSTON SCIENTIFIC - GALWAY
ballybrit business park
galway
Manufacturer Contact
linda leimer
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key4622791
MDR Text Key5762913
Report Number2134265-2015-01369
Device Sequence Number1
Product Code NIM
Combination Product (y/n)N
PMA/PMN Number
P050019
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 02/23/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/23/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/12/2018
Device Model NumberH7493915010240
Device Catalogue Number39150-1024
Device Lot Number17537344
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/23/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/17/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-