• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - LARGO HOMECHOICE PRO; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - LARGO HOMECHOICE PRO; SYSTEM, PERITONEAL, AUTOMATIC DELIVERY Back to Search Results
Catalog Number 5C8310
Device Problems Degraded (1153); Volume Accuracy Problem (1675)
Patient Problem No Patient Involvement (2645)
Event Date 02/27/2015
Event Type  malfunction  
Event Description
During evaluation of a returned homechoice device, a baxter technician determined the homechoice machine failed fluid volume accuracy testing.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
(b)(4).The homechoice device was returned and an evaluation was completed.An external/internal inspection was performed and no issues were noted.The homechoice device received a returned instrument testing evaluation (rite).This evaluation included functional and electrical testing of the device.The device passed the electrical test but failed the rite functional test due to volumetric accuracy.Review of the device logs did not verify the reported issue.Upon conclusion of the investigation, the cause of the failure was determined to be deteriorated piston foam and cracked door piston.A service history review was performed and revealed no issues that could have caused or contributed to the reported issue.The piston foam and door piston were scrapped.The device was sent to service.This issue is currently being addressed through a capa.Should additional relevant information become available, a follow-up report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOMECHOICE PRO
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE - LARGO
7511 114th ave. north
largo FL 33773
Manufacturer (Section G)
BAXTER HEALTHCARE - LARGO
7511 114th ave. north
largo FL 33773
Manufacturer Contact
kinga almasan
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key4644263
MDR Text Key16187256
Report Number1416980-2015-13735
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
PMA/PMN Number
K102936
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Type of Report Initial
Report Date 03/06/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/31/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number5C8310
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/27/2015
Date Manufacturer Received03/06/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-