The device was returned to omsc for investigation.The investigation confirmed that both the coil sheath and the wire fractured in about 35 mm from the distal end.The wire bended in the distal side, and the fracture surface of the coil showed that the load was applied from the inside of the coil towards the outside.It was restricted for the distal tip of the subject device to bend because fibrous foreign objects entangled around the distal tip.As the result of checking the manufacturing record of the same lot, nothing abnormal detected.Based on the confirmation of the subject device, olympus concluded that this event occurred due to the fracture of the coil sheath and the wire.The cause of the fracture was presumed as provided below.First, an excessive bending force was applied on the subject device during endoscopy, and the coil sheath and the wire buckled.Next, the coil sheath fractured because the doctor compulsorily spun the device while the fibrous foreign bodies adhered to the distal end.Then, the wire was ruptured because the doctor pulled the slider compulsorily.This report is being submitted as a medical device report in an abundance of caution.
|