• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER MEDICAL-KALAMAZOO UNKNOWN_MEDICAL_PRODUCT; BED, AC-POWERED ADJUSTABLE HOSPITAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER MEDICAL-KALAMAZOO UNKNOWN_MEDICAL_PRODUCT; BED, AC-POWERED ADJUSTABLE HOSPITAL Back to Search Results
Catalog Number UNK_MED
Device Problems Break (1069); Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problem Electric Shock (2554)
Event Date 03/12/2015
Event Type  Injury  
Event Description
It was reported via repair work order that the inlet filter was damaged and allegedly shocked the nurse.
 
Event Description
It was reported via repair work order that the inlet filter was damaged and allegedly shocked the nurse.
 
Manufacturer Narrative
Evaluation of the unit involved in the reported incident found that the inlet filter was damaged and the power cord was a non-stryker power cord with a damaged ground pin.Conversation with a member of the user facility's clinical team confirmed that a nurse was allegedly shocked when she touched the inlet filter while it was still attached to the power cord.The nurse was reportedly sent to the er for evaluation as a standard protocol but no injury was found.She did report to experience pain and was prescribed muscle relaxers.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN_MEDICAL_PRODUCT
Type of Device
BED, AC-POWERED ADJUSTABLE HOSPITAL
Manufacturer (Section D)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
Manufacturer (Section G)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
Manufacturer Contact
mary klaver
3800 east centre avenue
portage, MI 49002
2693292100
MDR Report Key4649704
MDR Text Key5615115
Report Number0001831750-2015-00157
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/13/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/01/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberUNK_MED
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received03/13/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-