Model Number 20
Device Problem Material Fragmentation (1261)
Patient Problem No Patient Involvement (2645)
Event Date 03/03/2015
Event Type  malfunction  
Event Description
The customer contacted physio-control to report that their device's therapy connector was broken and that a broken pin was lodged inside the therapy connector.As a result, defibrillation may not be possible.There was no patient use associated with the reported event.
 
Manufacturer Narrative
(b)(4): physio-control examined the customer's device and verified the reported issue.Physio then replaced the therapy connector assembly and completed unrelated repairs.After observing proper device operation through functional and performance testing the unit was returned to the customer for use.
 
Manufacturer Narrative
Physio-control examined the removed therapy connector assembly and determined that the cause of the reported issue was that a broken pin was lodged inside socket 6 (energy select).Physio confirmed that therapy was not possible due to the broken pin being inside socket 6.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) 20 DEFIBRILLATOR/MONITOR
Type of Device
DEFIBRILLATORS, AUTOMATIC, EXTERNAL
Manufacturer (Section D)
PHYSIO-CONTROL, INC
11811 willows road ne
po box 97006
redmond WA 98073 970
Manufacturer (Section G)
PHYSIO-CONTROL, INC
11811 willows road ne
po box 97006
redmond WA 98073 970
Manufacturer Contact
jason march
4258674000
MDR Report Key4652670
MDR Text Key5601153
Report Number3015876-2015-00392
Device Sequence Number1
Product Code MKJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K130454
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date 03/04/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/02/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number20
Device Catalogue Number3202487
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/20/2015
Is the Reporter a Health Professional? Yes
Device Age11 YR
Event Location Hospital
Date Manufacturer Received04/15/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/24/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-