ABBOTT LABORATORIES ARCHITECT CA 125 II; OC 125 DEFINED ANTIGEN Back to Search Results
Catalog Number 02K45-23
Device Problem High Test Results (2457)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/13/2015
Event Type  malfunction  
Event Description
The customer observed a falsely elevated ca125 result on the architect i2000sr analyzer.The following data was provided: sid (b)(6) initial 232.4, repeat 11.4 u/ml which matched clinical picture.There was no impact to patient management reported.
 
Manufacturer Narrative
An evaluation is in process.A follow up report will be submitted when the evaluation is complete.(b)(4).An evaluation is in process.
 
Manufacturer Narrative
Further investigation of the customer issue included a review of the complaint text, a search for similar complaints, a review of labeling and accuracy testing.No adverse trend was identified for the customer's issue.Labeling was reviewed and found to be adequate.The product was not available for return.Accuracy testing, using a retained kit and three levels of an internal panel with known concentrations, met all specifications.Based on all available information and abbott diagnostics' complaint investigation, the assay performed as intended and no product deficiency was identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCHITECT CA 125 II
Type of Device
OC 125 DEFINED ANTIGEN
Manufacturer (Section D)
ABBOTT LABORATORIES
100 abbott park road
abbott park IL 60064 350
Manufacturer (Section G)
ABBOTT LABORATORIES
100 abbott park road
abbott park IL 60064 350
Manufacturer Contact
noemi romero-kondos, rn bsn
100 abbott park road
dept. 09b9, lccp1-3
abbott park, IL 60064-3537
224667-512
MDR Report Key4662513
MDR Text Key20375227
Report Number1415939-2015-00006
Device Sequence Number1
Product Code LTK
Combination Product (y/n)N
Reporter Country CodeSP
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 03/17/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/06/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/07/2015
Device Catalogue Number02K45-23
Device Lot Number42527M500
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/23/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/01/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
ARCHITECT I2000SR ANALYZER; LN 03M74-02 SN (B)(4)
-
-