MAUDE Adverse Event Report: DENTSPLY PROFESSIONAL MIDWEST E MICRO ELECTRIC MOTOR DETACHABLE HOSE; CONTROLLER, FOOT, HANDPIECE AND CORD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DENTSPLY PROFESSIONAL MIDWEST E MICRO ELECTRIC MOTOR DETACHABLE HOSE; CONTROLLER, FOOT, HANDPIECE AND CORD Back to Search Results
Catalog Number 875050
Device Problem Overheating of Device (1437)
Patient Problem Burn(s) (1757)
Event Type  malfunction  
Event Description
In this event it was reported that a midwest e hose was crimped upon installation and caused the handpiece attachment to overheat and burn a patient's lip; no intervention was required.
 
Manufacturer Narrative
Since overheating of a handpiece could necessitate medical/surgical intervention to preclude permanent damage to a body structure or permanent impairment of a body function, this malfunction would be likely to cause/contribute to a serious injury should it recur.As such, this event meets the criteria for reportability per 21 cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MIDWEST E MICRO ELECTRIC MOTOR DETACHABLE HOSE
Type of Device
CONTROLLER, FOOT, HANDPIECE AND CORD
Manufacturer (Section D)
DENTSPLY PROFESSIONAL
des plaines IL
Manufacturer Contact
helen lewis
221 w. philadelphia st., ste. 60
susquehanna commerce center w.
york, PA 17401
7178457511
MDR Report Key4669530
MDR Text Key5732539
Report Number1419322-2015-00023
Device Sequence Number1
Product Code EBW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K090937
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Dentist
Type of Report Initial
Report Date 03/05/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/02/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number875050
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/05/2015
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-