Model Number 120804F
Device Problem Defective Component (2292)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/27/2015
Event Type  malfunction  
Event Description
It was reported that before use the catheter tip was broken.No patient complications were reported.
 
Manufacturer Narrative
The device evaluation is anticipated.However, the complaint could not be confirmed without the completion of the product evaluation.A review of the manufacturing records indicated that the product met specifications upon release.A supplemental report will be forthcoming once the device is received for examination.
 
Manufacturer Narrative
We received one (b)(4) catheter without the packaging for examination.A close visual examination found a tubing fracture located at the #4 inflation hole.The three other inflation holes are collapsed (oval shaped).The break site sections match and there are no missing sections.The balloon inflates and did not leak.Both windings appear to be in good condition.The rest of the catheter body was found to be in good condition.An investigation has been initiated to consider any potential manufacturing factors that may have contributed to this complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOGARTY ARTERIAL EMBOLECTOMY CATHETER
Type of Device
CATHETER, EMBOLECTOMY
Manufacturer (Section D)
EDWARDS LIFESCIENCES PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer (Section G)
EDWARDS LIFESCIENCES PR
state rd indus pk 402 km 1.4
anasco PR 00610
Manufacturer Contact
lynn selawski
1 edwards way
irvine, CA 92614
9497564386
MDR Report Key4689980
MDR Text Key5638745
Report Number2015691-2015-00834
Device Sequence Number1
Product Code DXE
Combination Product (y/n)N
Reporter Country CodeKS
PMA/PMN Number
PREAMENDMENT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/23/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/14/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/02/2016
Device Model Number120804F
Device Lot Number59618292
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received03/23/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/15/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-