• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RANDOX LABORATORIES, LTD. AMMONIA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RANDOX LABORATORIES, LTD. AMMONIA Back to Search Results
Catalog Number AM1054
Device Problem Visual Prompts will not Clear (2281)
Patient Problem No Patient Involvement (2645)
Event Date 01/19/2015
Event Type  malfunction  
Event Description
Patient sample will recover with a positive bias of approximately 60 umol/l when using ammonia catalogue number am1054 batch 326180.All affected customers have been contacted and instructed to discontinue use and discard, their inventory will be replaced.The ammonia test is generally ordered along with glucose, urea, electrolytes, kidney and liver functions tests alp, ast and alt therefore it is not considered a single conclusive diagnostic test to initiate therapy.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMMONIA
Type of Device
AMMONIA
Manufacturer (Section D)
RANDOX LABORATORIES, LTD.
55 diamond rd.
crumlin, county antrim
UK 
Manufacturer Contact
55 diamond rd.
crumlin, county antrim 
894422413
MDR Report Key4692040
MDR Text Key5672375
Report Number8020890-2015-00002
Device Sequence Number1
Product Code JIF
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K030873
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial
Report Date 02/02/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/13/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2016
Device Catalogue NumberAM1054
Device Lot Number326180
Is the Reporter a Health Professional? No
Date Manufacturer Received01/19/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/01/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-