The hospital reported that during an endoscopic vein harvesting procedure, first a regular vh-3200 was opened.When the harvester found tip altered the focus of the camera, she opened an accessory kit (vh-2004).That tip was found to be cloudy as well so a second accessory pack was opened.At that point, the sale representative was called into the case from the other room and had her insert a tip into the leg to see if that would make a difference.She could focus easier at that point and proceeded with the case.A replacement device was used to complete the procedure.The hospital did not report any patient effects.Refer to manufacture report number 2242352-2015-00278 and 2242352-2015-00293 for the other two related events.
|