• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH SECUR-FIT ADVANCED BROACH STRAIGHT SIZE 11; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, NON-POROUS, CALICU

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH SECUR-FIT ADVANCED BROACH STRAIGHT SIZE 11; PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, NON-POROUS, CALICU Back to Search Results
Catalog Number 1601-1011A
Device Problems Compatibility Problem (2960); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/24/2015
Event Type  malfunction  
Event Description
Surgeon stated that he believes there is a mismatch between the broaches and the implant.
 
Manufacturer Narrative
When completed, the evaluation summary will be submitted in a supplemental report.
 
Manufacturer Narrative
The event was not confirmed.A dimensional inspection was performed on the stem and found that it is within specifications.There is no indication at this time that the design, materials, or manufacturing of the subject device contributed to the event.If additional information becomes available, this investigation will be reopened.
 
Event Description
Surgeon stated that he believes there is a mismatch between the broaches and the implant.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SECUR-FIT ADVANCED BROACH STRAIGHT SIZE 11
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, UNCEMENTED, METAL/POLYMER, NON-POROUS, CALICU
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer Contact
william hanna
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key4694607
MDR Text Key15973049
Report Number0002249697-2015-01223
Device Sequence Number1
Product Code MEH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K122853
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative,company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/24/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/15/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number1601-1011A
Device Lot NumberMJPV2Y
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received09/24/2015
Was Device Evaluated by Manufacturer? No
Date Device Manufactured10/25/2010
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Other;
-
-