MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. BLADELESS TROCAR WITH STABILITY SLEEVE; LAPAROSCOPE, GENERAL AND PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. BLADELESS TROCAR WITH STABILITY SLEEVE; LAPAROSCOPE, GENERAL AND PLASTIC SURGERY Back to Search Results
Catalog Number B5LT
Device Problem Material Distortion (2977)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/25/2015
Event Type  malfunction  
Event Description
It was reported that during a lap sigmoidectomy, it was found that the tip of the device was distorted through a viewing endoscope.Another device was used to complete the case.There were no adverse consequences to the patient.
 
Manufacturer Narrative
(b)(4).Should the information be provided later, a supplemental medwatch will be sent.The analysis results found that only the sleeve assembly from a b5lt device was returned for analysis.As the obturator was not returned, we were unable to investigate further the reported incident; however the sleeve was observed to be damaged at the tip.Upon visual inspection with magnification, the sleeve was noted to be melted.See pictures.This type of damage is consistent with the one produced by an energized device.In order to prevent this kind of damage please avoid the contact with laser, electrosurgical or ultrasonic devices.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BLADELESS TROCAR WITH STABILITY SLEEVE
Type of Device
LAPAROSCOPE, GENERAL AND PLASTIC SURGERY
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo PR 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC
475 calle c
guaynabo PR 0096 9
Manufacturer Contact
guillermo villa
route 22 west po box 151
somerville, NJ 08876
9082180707
MDR Report Key4695295
MDR Text Key5640456
Report Number3005075853-2015-02362
Device Sequence Number1
Product Code GCJ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K032676
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 03/08/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/15/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberB5LT
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/31/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received04/14/2015
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-