(b)(4).Should the information be provided later, a supplemental medwatch will be sent.The analysis results found that only the sleeve assembly from a b5lt device was returned for analysis.As the obturator was not returned, we were unable to investigate further the reported incident; however the sleeve was observed to be damaged at the tip.Upon visual inspection with magnification, the sleeve was noted to be melted.See pictures.This type of damage is consistent with the one produced by an energized device.In order to prevent this kind of damage please avoid the contact with laser, electrosurgical or ultrasonic devices.
|