• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET ORTHOPEDICS COMP RVRS SHLDR GLNSP STD 36MM; PROSTHESIS, SHOULDER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET ORTHOPEDICS COMP RVRS SHLDR GLNSP STD 36MM; PROSTHESIS, SHOULDER Back to Search Results
Model Number N/A
Device Problem Positioning Problem (3009)
Patient Problem No Code Available (3191)
Event Date 03/19/2015
Event Type  Injury  
Event Description
It was reported that the patient had a reverse shoulder procedure on (b)(6) 2015.Subsequently, the patient was revised on (b)(6) 2015 due to the glenosphere not being fully seated into the baseplate.It is unknown why the glenosphere was not fully seated.The glenosphere and taper adaptor were removed and replaced.
 
Manufacturer Narrative
Current information is insufficient to permit a conclusion as to the cause of the event.Review of device history records show that lot released with no recorded anomaly or deviation.There are warnings in the package insert that state that this type of event can occur: "disassociations of modular components have been reported.Failure to properly align and completely seat the components together can lead to disassociation.Thoroughly clean and dry tapers prior to attachment of modular components to avoid crevice corrosion and improper seating.All additional locking screws must be adequately tightened.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COMP RVRS SHLDR GLNSP STD 36MM
Type of Device
PROSTHESIS, SHOULDER
Manufacturer (Section D)
BIOMET ORTHOPEDICS
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
BIOMET ORTHOPEDICS
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
megan haas
56 e. bell drive
warsaw, IN 46582
5743726700
MDR Report Key4703380
MDR Text Key5718390
Report Number0001825034-2015-01542
Device Sequence Number1
Product Code PAO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK080642
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/20/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/17/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Model NumberN/A
Device Catalogue Number115310
Device Lot Number996720
Other Device ID NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/20/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/03/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age67 YR
-
-