• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION EON MINI IPG, 16-CHANNEL RECHARGEABLE; SCS IPG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION EON MINI IPG, 16-CHANNEL RECHARGEABLE; SCS IPG Back to Search Results
Model Number 3788
Device Problems No Device Output (1435); Use of Device Problem (1670); Moisture or Humidity Problem (2986)
Patient Problem Inadequate Pain Relief (2388)
Event Date 12/31/2014
Event Type  Injury  
Event Description
Further follow up identified the ipg was explanted and replaced which resolved the issue.Reportedly, the patient has effective stimulation postoperatively.
 
Manufacturer Narrative
Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Event Description
It was reported the patient's ipg became inoperable due to not charging it for about 2 months.A sjm representative confirmed the communication issue between scs ipg and multiple external devices.As a result, surgical intervention will be taken at a later date to address the issue.
 
Manufacturer Narrative
(b)(4).Recall: 1627487-07262012-002-r.This ipg serial number was included in field advisories.Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EON MINI IPG, 16-CHANNEL RECHARGEABLE
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key4708159
MDR Text Key5890658
Report Number1627487-2015-20244
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Remedial Action Recall
Type of Report Initial,Followup
Report Date 04/29/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/20/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date08/28/2012
Device Model Number3788
Device Lot Number3175230
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received04/29/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/03/2010
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number1627487-12192011-003-R
Patient Sequence Number1
Treatment
MODEL 3186 & 3163, SCS LEAD; MODEL 1192, SCS ANCHOR
Patient Outcome(s) Other;
Patient Age91 YR
-
-