Model Number MP5100(TW)_
Device Problem Infusion or Flow Problem (2964)
Patient Problem Air Embolism (1697)
Event Date 03/17/2015
Event Type  Injury  
Event Description
It was reported that a patient experienced an air embolism while undergoing a percutaneous coronary intervention.An edwards dpt kit and edwards pressure tubing were connected to a medrad avanta auto injection system that was being used for contrast injection during the procedure.Approximately 3 to 4cc of air was confirmed on the monitor during imaging of the right coronary artery.The patient experienced chest pain and st elevations.Atropine and noradrenaline were given and the symptoms were resolved.There were no further patient complications reported.
 
Manufacturer Narrative
The lot number was not provided; therefore, a review of the manufacturing records could not be completed.Our lab received one dpt with attached terumo 20ml syringe at the female luer of dp t.The reported issue of air embolism issue was not able to be confirmed during evaluation.However, general evaluation was performed for the returned sample.As received into the laboratory, the connection between the dpt and terumo 20ml syringe appeared tight.No leakage was observed from the unit during leak test.The dpt zeroed and sensed pressure accurately on the pressure monitor.Electrical testing showed that dpt electronic components were intact because both input impedance and output impedance were within specifications.The zero-offset also met specifications.No visible damage/defect was found from returned unit during visual examination.There are no indications that this is related to the manufacturing process.No actions will be taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CUSTOM DEFINED PRODUCT
Type of Device
TRANSDUCER, PRESSURE, CATHETER TIP
Manufacturer (Section D)
EDWARDS LIFESCIENCES DR
carretera sanchez, km 18 1/2
parque ind. de itabo
haina. san cristobal
DR 
Manufacturer (Section G)
EDWARDS LIFESCIENCES DR
carretera sanchez, km 18 1/2
parque ind. de itabo
haina. san cristobal
DR  
Manufacturer Contact
lynn selawski
1 edwards way
irvine, CA 92614
9497564386
MDR Report Key4712174
MDR Text Key5740613
Report Number2015691-2015-00887
Device Sequence Number1
Product Code DXO
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Health Professional
Type of Report Initial
Report Date 03/24/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/21/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMP5100(TW)_
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/30/2015
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/24/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Weight40
-
-