It was reported to covidien on (b)(4) 2015 that a customer had an issue with an adapter.The customer states that the home patient went in to have her transfer set replaced because there was a hole in the catheter.The rn attempted to install the transfer set but noticed that it was not fitting well to the adapter.Upon connection, it did not look secure.The transfer set then fell off about a week later, (b)(6) 2015.The patient had been using peritoneal dialysis (pd) for about two years and was performing automated peritoneal dialysis (apd).There was no instrumentation or device used in an attempt to make the connection.Alcavis was used to sterilize the connection.There was no damage or residue on the threads of the connector noticed.The covidien adaptors looked perfectly normal.The rn was not aware of anything that may have caused the difficulty with the connection.The patient was the operator of the device.There was no patient injury, medical intervention, or adverse reaction reported with this event.The patient was given antibiotics as a precaution.There was a culture done, but the results were not known at the time of conversation.
|
The manufacturing lot number associated with this complaint was not provided.Without the lot number, a device history record (dhr) review could not be performed.All dhr's are reviewed for accuracy prior to product release.The complaint sample was not returned to the manufacturing site for review.Without the sample, it is not possible to determine a confirmed root cause of this issue.Should the sample be returned in the future, this complaint will be re-opened for further investigation.The instructions for use (ifu) indicates that infection is a potential complication inherent to these medical procedures and does not have association with the devices function.It must be noted that in-process controls (such as personnel training, incoming quality acceptance testing for raw material, 100% in process visual inspection and visual acceptance sampling) are in place to prevent nonconforming product from leaving the manufacturing operations.As per procedure, incoming performs acceptance sampling inspection, which would identify issues in the adapter ports.This complaint will be used for tracking and trending purposes.
|