STRYKER INSTRUMENTS-KALAMAZOO SMARTLIFE SMALL ASPTC HOUSING; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
|
Back to Search Results |
|
Catalog Number 7222120000 |
Device Problems
Fracture (1260); Device Disinfection Or Sterilization Issue (2909)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 04/01/2015 |
Event Type
malfunction
|
Event Description
|
It was reported that during testing conducted at the manufacturer facility, the latch on the battery housing was broken.As a result, the device would not lock closed, with the potential to expose a non-sterile battery to a surgical site.
|
|
Manufacturer Narrative
|
The reported event, housing will not lock as latch is broken, was duplicated.Service evaluation determined that the housing would not lock.The latch was broken on the bottom housing, which was physically cracked.The housing area where the latch was attached was broken off.As this is not a repairable device, it was not returned to the customer.
|
|
Event Description
|
It was reported that during testing conducted at the manufacturer facility, the latch on the battery housing was broken.As a result, the device would not lock closed, with the potential to expose a non-sterile battery to a surgical site.
|
|
Manufacturer Narrative
|
Failure analysis is in progress; additional information will be submitted once the quality investigation is completed.
|
|