Model Number H749CBM3250100
Device Problems Hole In Material (1293); Failure to Advance (2524)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Event Description
It was reported that a balloon pinhole occurred.The target lesion was located in the proximal right coronary artery (rca).A 10/2.50 flextome® cutting balloon® catheter was selected for use.During procedure, the device failed to cross the very tight target lesion.The device was then removed from the patient's body.It was then observed that there was blood in the shaft of the device.The device was flushed through the shaft; it was then noted that the balloon had a hole.There were no patient complications reported.
 
Manufacturer Narrative
(b)(6).Device evaluated by mfr: it is indicated that the device will not be returned for evaluation.A review of the batch history, historical trending, and similar complaint trending review for the product family will be conducted.If there is any further relevant information from that review, a supplemental medwatch will be filed.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXTOME® CUTTING BALLOON®
Type of Device
CATHETERS, TRANSLUMINAL CORONARY ANGIOPLASTY, PERCUTANEOUS
Manufacturer (Section D)
BOSTON SCIENTIFIC - GALWAY
ballybrit business park
galway
Manufacturer (Section G)
BOSTON SCIENTIFIC - GALWAY
ballybrit business park
galway
Manufacturer Contact
linda leimer
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key4740904
MDR Text Key5768820
Report Number2134265-2015-02672
Device Sequence Number1
Product Code LOX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P950020
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 04/13/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/30/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/07/2018
Device Model NumberH749CBM3250100
Device Catalogue NumberCBM325010
Device Lot Number17584284
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/13/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/29/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age77 YR
-
-