A site representative, radiologic technologist (rt) reported that during a spinal fusion procedure, the imaging system detector positioner cover was damaged.Issue noticed when opening the o-arm door to remove it from around the patient after the post-op 3d spin.A popping noise was heard when opening the door.The surgeon completed the procedure with the use of the imaging system.There was no impact on patient outcome.
|
Patient weight not available from the site.A medtronic representative, following up with the site, tested the system and reported both louver hinges were broken from sterile drapes caught inside the door when 3d spin was acquired.Per imaging system user manual: "caution: carefully watch while the door closes to make sure that no part of the drape is caught inside the door.This could cause internal failure of the components." replacement hinges shipped to site for issue resolution.On (b)(4) 2015, the medtronic representative replaced the hinges and performed an imaging system check-out, all areas passed.The system was reported fully functional after the replacement of the hinges.No parts have been received by the manufacturer for analysis.
|