• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LITTLETON) O-ARM 1000 IMAGING SYSTEM; IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LITTLETON) O-ARM 1000 IMAGING SYSTEM; IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE Back to Search Results
Model Number BI-700-00027-120
Device Problems Device Issue (2379); Difficult to Open or Close (2921)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/08/2015
Event Type  malfunction  
Event Description
A site representative, radiologic technologist (rt) reported that during a spinal fusion procedure, the imaging system detector positioner cover was damaged.Issue noticed when opening the o-arm door to remove it from around the patient after the post-op 3d spin.A popping noise was heard when opening the door.The surgeon completed the procedure with the use of the imaging system.There was no impact on patient outcome.
 
Manufacturer Narrative
Patient weight not available from the site.A medtronic representative, following up with the site, tested the system and reported both louver hinges were broken from sterile drapes caught inside the door when 3d spin was acquired.Per imaging system user manual: "caution: carefully watch while the door closes to make sure that no part of the drape is caught inside the door.This could cause internal failure of the components." replacement hinges shipped to site for issue resolution.On (b)(4) 2015, the medtronic representative replaced the hinges and performed an imaging system check-out, all areas passed.The system was reported fully functional after the replacement of the hinges.No parts have been received by the manufacturer for analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
O-ARM 1000 IMAGING SYSTEM
Type of Device
IMAGE-INTENSIFIED FLUOROSCOPIC X-RAY SYSTEM, MOBILE
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LITTLETON)
300 foster street
littleton MA 01460
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC.
826 coal creek circle
louisville CO 80027 971
Manufacturer Contact
kristianna bilan
826 coal creek circle
louisville, CO 80027-9710
7208902338
MDR Report Key4743670
MDR Text Key15899558
Report Number1723170-2015-00542
Device Sequence Number1
Product Code OXO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K050996
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Radiologic Technologist
Type of Report Initial
Report Date 04/08/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/01/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Radiologic Technologist
Device Model NumberBI-700-00027-120
Device Catalogue Number9732719
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/08/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/14/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age45 YR
-
-