MEDTRONIC PUERTO RICO OPERATIONS CO. SYNCHROMED II; PUMP, INFUSION, IMPLANTED, PROGRAMMABLE
|
Back to Search Results |
|
Model Number 8637-40 |
Device Problem
Torn Material (3024)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Type
Injury
|
Event Description
|
It was reported that the patient was having surgery to suture down her pump because it kept flipping.The pump flipping did not cause any patient symptoms.The pump was flipping due to movement in the pocket; the movement was attributed to broken suture loops.The patient was doing well, was receiving effective therapy, and getting good pain relief.The patient status at the time of the report was alive with no injury.The pump system was being used to infuse fentanyl and bupivacaine.
|
|
Manufacturer Narrative
|
Concomitant products: product id: 8709, lot# l78422, implanted: (b)(6) 2000, product type: catheter.Product id: 8578, serial# (b)(4), implanted: (b)(6) 2014, product type: accessory.(b)(4).
|
|