• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES USA HANDLE FOR 90° SCREWDRIVER; DRILL, BONE, POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES USA HANDLE FOR 90° SCREWDRIVER; DRILL, BONE, POWERED Back to Search Results
Catalog Number 03.505.004
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/14/2015
Event Type  malfunction  
Event Description
Device report from synthes (b)(4) reports an event in (b)(6) as follows: it was reported that the surgeon experienced difficulty using the contra-angle hand piece attachments during a sagittal split ramus osteotomy (ssro) procedure on (b)(6) 2015.While drilling with the hand piece in high speed, the surgeon noted that it would work and then stop.The surgeon was able to drill while the device was working and finished the surgery without delay.During a postoperative check of the device, a rust-like, brown substance was discovered on the mortar of the inside shaft tube.This report is 2 of 2 for (b)(4).
 
Manufacturer Narrative
Additional narrative: patient information is not available for reporting.Additional product codes for this report include dzj.Device is an instrument and is not implanted or explanted.Complainant part is expected to be returned for manufacturer review/investigation, but has yet to be received.Investigation could not be completed and no conclusion could be drawn as no device was returned.Without a lot number, the device history record review could not be requested.Device used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HANDLE FOR 90° SCREWDRIVER
Type of Device
DRILL, BONE, POWERED
Manufacturer (Section D)
SYNTHES USA
1302 wrights lane east
west chester PA 19380
Manufacturer (Section G)
SYNTHES USA
Manufacturer Contact
linda plews
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key4747432
MDR Text Key22152042
Report Number2520274-2015-13585
Device Sequence Number1
Product Code DZI
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
PK082649
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Health Professional
Type of Report Initial
Report Date 04/16/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/05/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number03.505.004
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/16/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-