Review of historical complaint trending and lot search data for architect ca 125 ii reagent lot 41540m500 did not identify an increase in complaint activity for the issue under investigation.Accuracy testing was completed using retained kits of reagent lot 41540m500.Acceptance criteria was met, which indicates acceptable product performance.A deficiency was not identified as panel testing shows that the likely cause lot performs per specification.No malfunction was identified since the issue involves a discreet sample.Per the complaint text, the sample was re-centrifuged and retests on two different instruments produced the expected results.Additionally, the specimen collection and preparation for analysis section of the architect ca 125 ii package insert provides conditions that may have contributed to the erroneous result.
|