A steris service technician inspected the table and found that multiple table top sections were missing x-ray top metal grommets.The technician further noted that the slide position on the hand control was displaying the position inaccurately and fluctuating.During the time of the reported event the surgical table had x-ray tops installed.The technician replaced the slide sensor, tested the table and confirmed it to be operating properly.No additional issues have been reported.The user facility is in the process of ordering new table tops.Once the table tops are received the user facility will replace the existing table tops and x-ray top metal grommets.The table is not under steris service contract and is service and maintained by the user facility.The operator manual states (pp.5-7), "remove and inspect x-ray tops on each table section (if equipped) 1 x per year." in addition, the operator manual states (pp.2-3), "patient positioning and weight limitation- note: when positioning the patient on the table, note the following: 4) ensure all accessories are properly installed and secured.".
|