• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. ECHELON; INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. ECHELON; INSTRUMENT Back to Search Results
Model Number OR71364006
Device Problem Fracture (1260)
Patient Problem No Code Available (3191)
Event Date 04/16/2015
Event Type  Injury  
Event Description
It was reported that surgery was delayed due to the instrument fracturing during surgery.Back up product was used to complete the surgery.
 
Manufacturer Narrative
 
Manufacturer Narrative
The associated t handle was returned and evaluated.Our investigation included a visual inspection of the device which showed that the jaws of the t handle have fractured in two places.The fractured pieces were not returned.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.The device was manufactured in 2013.The t-handle fractured in the jaws of the handle, likely from static bending or torsional overload.An overload fracture can occur if the mechanical loads applied to the device exceed the strength of the material.Our investigation did not determine a specific cause of the stated failure.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This investigation could not verify or identify any evidence of product contribution to the reported problem.Based on this investigation, the need for corrective action is not indicated.We recommend that all re-usable instruments be routinely inspected for wear/damage and replaced as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ECHELON
Type of Device
INSTRUMENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
Manufacturer (Section G)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
Manufacturer Contact
melanie travis
1450 brooks road
memphis, TN 38116
9013996233
MDR Report Key4759386
MDR Text Key5953566
Report Number1020279-2015-00261
Device Sequence Number1
Product Code HAO
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative,company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/16/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/08/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Other
Device Model NumberOR71364006
Device Catalogue NumberOR71364006
Device Lot Number13GSY0014
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/20/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received04/16/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
-
-