Model Number H802220200391
Device Problems Loose or Intermittent Connection (1371); Device Operates Differently Than Expected (2913)
Patient Problem No Patient Involvement (2645)
Event Date 04/14/2015
Event Type  malfunction  
Event Description
It was reported that the turbine pressure control knob cannot control the speed.A rotablator rotational atherectomy system were selected for use.During preparation the rotational speed was tested and it was then noted that the high speed and low speed rate were the same.The dynaglide displayed 160,000rpm.It was also noted that the turbine pressure control knob was loosened and it cannot adjust the speed.The procedure was completed with another of the same device.There were no patient complications reported and the patient's condition was stable.
 
Manufacturer Narrative
(b)(4).Device evaluated by mfr: it is indicated that the device will not be returned for evaluation.A review of the batch history, historical trending, and similar complaint trending review for the product family will be conducted.If there is any further relevant information from that review, a supplemental medwatch will be filed.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTABLATOR ROTATIONAL ATHERECTOMY SYSTEM
Type of Device
CATHETER, CORONARY, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC - FREMONT (CE)
47215 lakeview blvd
north dock
fremont CA 94538
Manufacturer (Section G)
BOSTON SCIENTIFIC - FREMONT (CE)
47215 lakeview blvd
north dock
fremont CA 94538
Manufacturer Contact
linda leimer
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key4762935
MDR Text Key5780748
Report Number2134265-2015-02826
Device Sequence Number1
Product Code MCX
Combination Product (y/n)N
Reporter Country CodeTW
PMA/PMN Number
P900056
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 04/14/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/11/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberH802220200391
Device Catalogue Number22020-039
Device Lot NumberRC108079
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/14/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/02/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-