• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: HILL-ROM, INC. VERSACARE BED; A/C POWERED ADJUSTABLE HOSPITAL BED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HILL-ROM, INC. VERSACARE BED; A/C POWERED ADJUSTABLE HOSPITAL BED Back to Search Results
Model Number 3200
Device Problem Device Inoperable (1663)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 04/14/2015
Event Type  malfunction  
Event Description
Hill-rom received a report from the account stating the brakes were not working.The bed was located in 1 rehab at the account.There was no patient/user injury reported.This report was filed in our complaint handling system as complaint #(b)(4).
 
Manufacturer Narrative
The technician found the left foot casters was the issue.A search of the hill-rom maintenance records showed hill-rom performed preventative maintenance on this bed in 2014.It is unknown if the facility performed any other preventative maintenance on this bed.The technician replaced the left foot brake caster to resolve the issue.Based on this information, no further action is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VERSACARE BED
Type of Device
A/C POWERED ADJUSTABLE HOSPITAL BED
Manufacturer (Section D)
HILL-ROM, INC.
batesville IN 47006
Manufacturer Contact
tony werner
1069 state route 46 east
batesville, IN 47006
8129312359
MDR Report Key4768296
MDR Text Key20725988
Report Number1824206-2015-00576
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 04/14/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/08/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number3200
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received04/14/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/01/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-