Procedure type: nephrectomy.According to the reporter: while using the stapling device articulated to the users left, the stapler became stuck in that position.It would not articulate, close, or open.They had to manipulate the stapler out of the patient through the trocar.Once the device was out of the patient, they tried to remove the reload.The device began to go through the recycling/starting sequence and then articulated the reload again to the users left.The device was able to be removed.No reinforcement material was used.There was no unanticipated tissue loss.The incision was not extended by more than one inch.There was no unanticipated blood loss of 500cc or more.Surgery time was not delayed by more than 30 minutes.No device fragment fell into the patient.There were no adverse effects on the patient due to this incident.
|