Model Number MCS-P3-29-AOA-US
Device Problems Patient-Device Incompatibility (2682); Device-Device Incompatibility (2919); Device Dislodged or Dislocated (2923)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/27/2015
Event Type  Injury  
Event Description
Medtronic received information that during the implant of this transcatheter bioprosthetic valve, the valve dislodged into the sinuses.It was reported the dislodgement was due to large patient anatomy with large vessels and the presence of a previously implanted non-medtronic surgical valve.This dislodged valve was snared to just above the sinotubular junction.A second valve was implanted successfully.No adverse patient effects were reported.
 
Manufacturer Narrative
The device remains implanted, therefore no product analysis can be performed.Without return of the product, no definitive conclusions can be drawn regarding the clinical observation.Should the device be returned or additional information become available, a supplemental report will be submitted.(b)(4).
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COREVALVE 29MM AORTIC VALVE
Type of Device
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED
Manufacturer (Section D)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
Manufacturer (Section G)
MEDTRONIC STRUCTURAL HEART
8200 coral sea street ne
mounds view MN 55112
Manufacturer Contact
paula bixby
8200 coral sea street ne
mounds view, MN 55112
7635055378
MDR Report Key4783190
MDR Text Key20151300
Report Number2025587-2015-00570
Device Sequence Number1
Product Code NPT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P130021
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative,company representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/27/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/19/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date02/22/2017
Device Model NumberMCS-P3-29-AOA-US
Device Catalogue NumberMCS-P3-29-AOA-US
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/27/2015
Was Device Evaluated by Manufacturer? No
Date Device Manufactured03/05/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;