The testing performed indicates that the injector problem may have occurred as the flap fixing is missing from the injector.This damage is likely to have occurred during manufacture; however, it is not possible to confirm at what stage of the manufacturing process the damage may have occurred.A review of existing vigilance data from the month of manufacture of the superflex aspheric 920h iol (september 2013) was carried out in order to determine if any trends existed.This review confirmed that no other incidents, of any type, have been received against the superflex aspheric 920h iol batch 093e51753 or the injector batch associated with this case report.
|