• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION EON MINI IPG, 16-CHANNEL RECHARGEABLE; SCS IPG

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION EON MINI IPG, 16-CHANNEL RECHARGEABLE; SCS IPG Back to Search Results
Model Number 3788
Device Problems Nonstandard Device (1420); Communication or Transmission Problem (2896); Unintended Movement (3026)
Patient Problems Inadequate Pain Relief (2388); Weight Changes (2607)
Event Type  Injury  
Manufacturer Narrative
Results: the complaint was confirmed.As returned, the ipg would not communicate due to a depleted battery.The battery was recovered and the ipg communicated, charged, and was tested to manufacturing specifications using the autotester.The ipg passed all tests.Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Event Description
It was reported the patient lost stimulation approximately a month ago.In addition, the ipg will no longer communicate with the charger.Follow-up information identified the ipg had flipped in the pocket due to the patient 's weight loss.Surgical intervention was undertaken on (b)(6) 2015 and the ipg was explanted and replaced with a different model.Stimulation was restored postoperatively thus resolving the issue.
 
Manufacturer Narrative
Udi(di) (b)(4).This ipg correction number was included in field advisories.1627487-05242011-002-r, 1627487-07262012-002-r, 1627487-12192011-003-r.Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EON MINI IPG, 16-CHANNEL RECHARGEABLE
Type of Device
SCS IPG
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key4795631
MDR Text Key14378653
Report Number1627487-2015-08299
Device Sequence Number1
Product Code GZB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010032
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Remedial Action Recall
Type of Report Initial,Followup
Report Date 07/06/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/26/2015
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/28/2011
Device Model Number3788
Device Lot Number2865284
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/21/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received07/06/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/28/2009
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
MODEL 3214, SCS LEAD
Patient Outcome(s) Other;
-
-