Device Problems
Nonstandard Device (1420); Communication or Transmission Problem (2896); Unintended Movement (3026)
|
Patient Problems
Inadequate Pain Relief (2388); Weight Changes (2607)
|
Event Type
Injury
|
Manufacturer Narrative
|
Results: the complaint was confirmed.As returned, the ipg would not communicate due to a depleted battery.The battery was recovered and the ipg communicated, charged, and was tested to manufacturing specifications using the autotester.The ipg passed all tests.Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
|
|
Event Description
|
It was reported the patient lost stimulation approximately a month ago.In addition, the ipg will no longer communicate with the charger.Follow-up information identified the ipg had flipped in the pocket due to the patient 's weight loss.Surgical intervention was undertaken on (b)(6) 2015 and the ipg was explanted and replaced with a different model.Stimulation was restored postoperatively thus resolving the issue.
|
|
Manufacturer Narrative
|
Udi(di) (b)(4).This ipg correction number was included in field advisories.1627487-05242011-002-r, 1627487-07262012-002-r, 1627487-12192011-003-r.Sjm has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Sjm defers to the patient's physician regarding medical history.
|
|