Model Number 1600
Device Problem Device Inoperable (1663)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 04/24/2015
Event Type  malfunction  
Event Description
Hill-rom received a report from the account stating nurse call was not working.The bed was located at the account.There was no patient/user injury reported.(b)(4).
 
Manufacturer Narrative
The hill-rom technician found the communication cable was the faulty.A search of the hill-rom maintenance records showed hill-rom performed preventative maintenance on this bed in 2014.It is unknown if the facility performed any other preventative maintenance on this bed.The technician replaced the communication cable to resolve the issue.Based on this information, no further action is required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANTA BED
Type of Device
A/C POWERED ADJUSTABLE HOSPITAL BED
Manufacturer (Section D)
HILL-ROM, INC.
batesville IN
Manufacturer Contact
john cummings
1069 state rte 46 e
batesville, IN 47006
8129312869
MDR Report Key4798680
MDR Text Key17695601
Report Number1824206-2015-00599
Device Sequence Number1
Product Code FNL
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Unknown
Type of Report Initial
Report Date 04/24/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/20/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number1600
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received04/24/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/01/2002
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-