Reportable based on device analysis completed on 06may2015.It was reported that the device failed to deploy.The stenosed target lesion was located in the left carotid artery.A 190cm filterwire ez¿ was selected and advanced to treat the target lesion.During procedure, after the device reached the lesion, it was noted that the device would not deploy.There were no patient complications reported and the patient's status was stable.However, device analysis revealed the spring tip is broken and missing.
|
(b)(4).The device was returned for analysis.The unit returned has the distal tip damaged.Visual inspection observed that the wire was found exposed through sheath slit, therefore, sheathing and unsheathing test could not be performed.Moreover, the wire is bent and the spring tip was partially broken (coilwire and corewire).The filter bag was observed to be in good condition and met specification.All the outer diameter (ods) taken and all of them are according specifications.The most distal section of the spring tip is broken and missing.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is operational context as device performance was limited due to anatomical procedural factors.(b)(4).
|