(b)(4).There was no required intervention to prevent permanent impairment/damage in this case.Evaluation summary: one delivery system was returned to medtronic for evaluation.The capsule was not returned for evaluation.The delivery system was investigated visually for external damage.There were no signs of blood or tissue on the device.The delivery system was not bent; the plunger was not broken, but the plunger was not pushed to the end.The emergency release on the delivery system was not implemented.The capsule electrodes were visually acceptable.The wire that holds the capsule was completely off and the foam gasket was in good condition.The delivery system did not have any visible damage.As the product was received, the device functioned per specification.The plunger not pushed to the end is indicative of mishandling.
|