Catalog Number 5C8310R
Device Problem Inappropriate/Inadequate Shock/Stimulation (1574)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/11/2015
Event Type  malfunction  
Event Description
It was reported that the patient was shocked by the homechoice device when they went to reload a new set.The patient turned the device off and unplugged it.The technical service representative discussed the use of manual supplies with the patient.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
(b)(4).The device has been received and the evaluation is in progress.Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
(b)(4).The device was received and the evaluation is completed.The event history log review showed no keystrokes, programming, or use related events that indicated and/or contributed to the reported issue.A service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.An internal and external visual inspection was performed.The homechoice device received a returned instrument testing evaluation (rite).This evaluation includes functional and electrical testing on the device.The reported condition was not verified.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HOMECHOICE PRO
Type of Device
SYSTEM, PERITONEAL, AUTOMATIC DELIVERY
Manufacturer (Section D)
BAXTER HEALTHCARE - SINGAPORE
2 woodlands industrial park
singapore 2573
SN  2573
Manufacturer (Section G)
BAXTER HEALTHCARE - SINGAPORE
2 woodlands industrial park
singapore 2573
SN   2573
Manufacturer Contact
kinga almasan
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key4802460
MDR Text Key5917506
Report Number1416980-2015-23176
Device Sequence Number1
Product Code FKX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102936
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/11/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/28/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Catalogue Number5C8310R
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/19/2015
Is the Reporter a Health Professional? No
Date Manufacturer Received06/15/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Age62 YR
-
-