AV-TEMECULA-CT ARMADA 14 PTA CATHETER; PERIPHERAL DILATATION CATHETER Back to Search Results
Catalog Number A2030-200
Device Problems Difficult to Remove (1528); Folded (2630)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 05/05/2015
Event Type  malfunction  
Event Description
It was reported that the procedure was to treat the anterior and posterior tibial.The armada 14 3 x 200 x150 mm was inflated at least 4 times to nominal pressure.When it was removed from the patient, it was very difficult to pull it back inside of the sheath and it appeared that the balloon was "bunching up".There was no clinically significant delay in the procedure or adverse patient effects.No additional information was provided.
 
Manufacturer Narrative
(b)(4).It is indicated that the device is not returning for evaluation; therefore, a failure analysis of the complaint device could not be completed.The lot history record (lhr) for this product could not be reviewed and a similar incident query could not be performed because the product was not returned for evaluation and the lot number was not reported.Based on the information reviewed, there is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Manufacturer Narrative
(b)(4).The customer reported the device was discarded.Investigation is not yet complete.A follow-up will be submitted with all relevant information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARMADA 14 PTA CATHETER
Type of Device
PERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 462
Manufacturer (Section G)
ABBOTT VASCULAR, TEMECULA, CA USA REG# 2024168
abbott vascular
26531 ynez road
temecula CA 92591 462
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key4805347
MDR Text Key5892595
Report Number2024168-2015-03017
Device Sequence Number1
Product Code LIT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102705
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 05/05/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/29/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberA2030-200
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/01/2015
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-