• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: TERUMO CARDIOVASCULAR SYSTEMS CORP. TERUMO TCM II COOLING AND HEATING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

TERUMO CARDIOVASCULAR SYSTEMS CORP. TERUMO TCM II COOLING AND HEATING SYSTEM Back to Search Results
Model Number 4415
Device Problems Device Reprocessing Problem (1091); Restricted Flow rate (1248)
Patient Problem No Patient Involvement (2645)
Event Date 05/07/2015
Event Type  malfunction  
Event Description
It was reported by the field svc rep (fsr) that during troubleshooting of mdr #1828100-2015-00437, the flow over the cold plate seemed less than normal on the cooler heater unit.There was no pt involvement.
 
Manufacturer Narrative
This complaint is related to mdr #1828100-2015-00161 and mrd #1828100-2015-00437.The field svc rep (fsr) corrected the low flow by cleaning out the internal filter.Flow now appeared normal and the cooler heater unit was operating in all modes.The unit operated to mfr spec and was returned to clinical use.If add'l info becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TERUMO TCM II COOLING AND HEATING SYSTEM
Type of Device
TCM II COOLING AND HEATING SYSTEM
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORP.
ann arbor MI 48103
Manufacturer Contact
jan winder
6200 jackson rd.
ann arbor, MI 48103
7346634145
MDR Report Key4808445
MDR Text Key5893185
Report Number1828100-2015-00438
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K883603
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Not Applicable
Type of Report Initial
Report Date 05/07/2015
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/28/2015
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number4415
Device Catalogue Number4415
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received05/07/2015
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-